Title: A Simple High-performance Liquid Chromatography Coupled to Fluorescence Detection Method using Column-switching Technique for Measuring Urinary 1-hydroxypyrene from Environmental Exposure
Authors: Kunrunya Sutan
Warangakana Naksen
Tippawan Prapamontol
Authors: Kunrunya Sutan
Warangakana Naksen
Tippawan Prapamontol
Issue Date: 2017
Publisher: Science Faculty of Chiang Mai University
Abstract: A method for determination of urinary 1-hydroxypyrene (1-OHP) was developed using HPLC-FLD with a column-switching system. 1-OHP in urine samples was extracted by b-glucuronidase hydrolysis, followed by C18 SPE with optimized conditions. LOD and LOQ were 0.01 and 0.03 ng/mL, respectively; this was sensitive enough to determine urinary 1-OHP in low-dose exposure. The precision and accuracy of four concentrations (0.06, 2.0, 8.0 and 25 ng/mL) were tested. The intra- and inter-batch precisions were between 1.82-4.22 and 5.01-15.8 %RSD, respectively, while accuracies were between 92.8-117% recoveries, which were in the acceptable range. This method was applied in 130 school children living in a seasonal forest-fire burning area in Northern Thailand. Urinary 1-OHP was detected in all urine samples at concentrations of 1.35±2.26 mmol/mol creatinine, and a range of 0.03-16.0 mmol/mol creatinine. The developed method offers a potential application to determine urinary 1-OHP levels from environmentally low exposure.
URI: http://it.science.cmu.ac.th/ejournal/dl.php?journal_id=8471
http://cmuir.cmu.ac.th/jspui/handle/6653943832/63971
ISSN: 0125-2526
Appears in Collections:CMUL: Journal Articles

Files in This Item:
There are no files associated with this item.


Items in CMUIR are protected by copyright, with all rights reserved, unless otherwise indicated.