Title: | Simple sequential injection analysis systems with a dynamic surface tension detector |
Authors: | Narong Lenghor Jaroon Jakmunee Bryan J. Prazen Robert E. Synovec Gary D. Christian Kate Grudpan |
Authors: | Narong Lenghor Jaroon Jakmunee Bryan J. Prazen Robert E. Synovec Gary D. Christian Kate Grudpan |
Keywords: | Chemistry |
Issue Date: | 1-Jan-2006 |
Abstract: | Simple sequential injection analysis systems with DSTD (SIA/DSTD) have been developed. One was employed for the study of the effects of the ion contents in solutions to the dynamic surface pressure of ionic surfactants. The results from the studies show the possibility for an alternative simple fast screening, but also a sensitive procedure for water quality determination. Another simple SIA/DSTD system has been demonstrated for the quantification of an anionic surfactant using a single standard calibration. 2006 © The Japan Society for Analytical Chemistry. |