Please use this identifier to cite or link to this item: http://cmuir.cmu.ac.th/jspui/handle/6653943832/49899
Full metadata record
DC FieldValueLanguage
dc.contributor.authorEduardo Yugo Suzukien_US
dc.contributor.authorBoonsiva Suzukien_US
dc.date.accessioned2018-09-04T04:19:59Z-
dc.date.available2018-09-04T04:19:59Z-
dc.date.issued2011-05-01en_US
dc.identifier.issn08895406en_US
dc.identifier.other2-s2.0-79955591942en_US
dc.identifier.other10.1016/j.ajodo.2010.11.017en_US
dc.identifier.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=79955591942&origin=inwarden_US
dc.identifier.urihttp://cmuir.cmu.ac.th/jspui/handle/6653943832/49899-
dc.description.abstractIntroduction: The purpose of this study was to analyze the maximum insertion torque (MIT) and maximum removal torque (MRT) values of orthodontic miniscrews. Methods: Two hundred eighty titanium miniscrews were placed in several sites in the maxillae and mandibles of orthodontic patients to provide skeletal anchorage. Two types of miniscrews were used: predrilling, cylindrical miniscrews, and self-drilling, conical miniscrews. A force of 50 g was applied with nickel-titanium coil springs 2 weeks after placement. MIT and MRT values were assessed with a digital torque gauge. Torque values were subjected to the paired t test, Student t test, chi-square test, and 1-way analysis of variance (ANOVA) (P <0.05). Results: MIT values were significantly higher for the self-drilling miniscrews (mean, 14.5 Ncm; 95% CI, 13.6-15.3) than for the predrilling miniscrews (mean, 9.2 Ncm; 95% CI, 8.6-9.9) in all implant sites. For both predrilling and self-drilling miniscrews, the highest MIT values were observed at the midpalatal suture site followed by the dentoalveolar bones of the mandible and maxilla, respectively. In contrast, MRT values were significantly higher for the predrilling miniscrews (mean, 22.6 Ncm; 95% CI, 22.0-23.3) than for the self-drilling miniscrews (mean, 17.6 Ncm; 95% CI, 16.8-18.4). Accordingly, the mean torque ratio of predrilling miniscrews was twice that of the self-drilling miniscrews. Conclusions: An inverse relationship between MIT and MRT values was observed. The results suggest that relatively lower MIT values were more favorable to osseointegration than higher values. Copyright © 2011 by the American Association of Orthodontists.en_US
dc.subjectDentistryen_US
dc.titlePlacement and removal torque values of orthodontic miniscrew implantsen_US
dc.typeJournalen_US
article.title.sourcetitleAmerican Journal of Orthodontics and Dentofacial Orthopedicsen_US
article.volume139en_US
article.stream.affiliationsChiang Mai Universityen_US
Appears in Collections:CMUL: Journal Articles

Files in This Item:
There are no files associated with this item.


Items in CMUIR are protected by copyright, with all rights reserved, unless otherwise indicated.