Please use this identifier to cite or link to this item: http://cmuir.cmu.ac.th/jspui/handle/6653943832/75825
Title: Flow injection system with amperometric detection for determination of iodine value of vegetable oils
Authors: Thidarat Kruatian
Kritsana Jitmanee
Jaroon Jakmunee
Authors: Thidarat Kruatian
Kritsana Jitmanee
Jaroon Jakmunee
Keywords: Biochemistry, Genetics and Molecular Biology;Chemistry;Materials Science;Mathematics;Physics and Astronomy
Issue Date: 1-Jan-2021
Abstract: A flow injection (FI) system with amperometric detection has been proposed for the determination of iodine value (IV) of vegetable oils. An aliquot of a sample of 0.010-g was treated with 1.00 mL of Hanus solution for 1 minute before adding with potassium iodide solution to generate tri-iodide. Then, the solution was injected into the FI-amperometric system to detect the tri-iodide by using a screen-printed carbon electrode (SPCE) at an applied potential of 750 mV (vs Ag/AgCl). The system was optimized to provide high sensitivity by utilizing various concentration of Hanus solutions for constructing a calibration graph, which was linear in a range of 40-100 µM (r2 = 0.997) with limit of detection of 3.6 IV. An injection frequency of 60 h-1 was achieved by using 2.5 mL/min flow rate. A good repeatability of injection (0.20-0.58 %RSD, n=11) and good reproducibility for replicated analysis of 5 aliquots of a sample (1.07-1.36 %RSD) were demonstrated. The proposed method was successfully applied to vegetable oil samples with low to high iodine value. The proposed method offers high sample throughput and low reagent consumption.
URI: https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85104957900&origin=inward
http://cmuir.cmu.ac.th/jspui/handle/6653943832/75825
ISSN: 01252526
Appears in Collections:CMUL: Journal Articles

Files in This Item:
There are no files associated with this item.


Items in CMUIR are protected by copyright, with all rights reserved, unless otherwise indicated.