Please use this identifier to cite or link to this item: http://cmuir.cmu.ac.th/jspui/handle/6653943832/56669
Title: Evaluation of intensity drift correction strategies using MetaboDrift, a normalization tool for multi-batch metabolomics data
Authors: Chanisa Thonusin
Heidi B. IglayReger
Tanu Soni
Amy E. Rothberg
Charles F. Burant
Charles R. Evans
Authors: Chanisa Thonusin
Heidi B. IglayReger
Tanu Soni
Amy E. Rothberg
Charles F. Burant
Charles R. Evans
Keywords: Biochemistry, Genetics and Molecular Biology;Chemistry;Medicine
Issue Date: 10-Nov-2017
Abstract: © 2017 Elsevier B.V. In recent years, mass spectrometry-based metabolomics has increasingly been applied to large-scale epidemiological studies of human subjects. However, the successful use of metabolomics in this context is subject to the challenge of detecting biologically significant effects despite substantial intensity drift that often occurs when data are acquired over a long period or in multiple batches. Numerous computational strategies and software tools have been developed to aid in correcting for intensity drift in metabolomics data, but most of these techniques are implemented using command-line driven software and custom scripts which are not accessible to all end users of metabolomics data. Further, it has not yet become routine practice to assess the quantitative accuracy of drift correction against techniques which enable true absolute quantitation such as isotope dilution mass spectrometry. We developed an Excel-based tool, MetaboDrift, to visually evaluate and correct for intensity drift in a multi-batch liquid chromatography – mass spectrometry (LC–MS) metabolomics dataset. The tool enables drift correction based on either quality control (QC) samples analyzed throughout the batches or using QC-sample independent methods. We applied MetaboDrift to an original set of clinical metabolomics data from a mixed-meal tolerance test (MMTT). The performance of the method was evaluated for multiple classes of metabolites by comparison with normalization using isotope-labeled internal standards. QC sample-based intensity drift correction significantly improved correlation with IS-normalized data, and resulted in detection of additional metabolites with significant physiological response to the MMTT. The relative merits of different QC-sample curve fitting strategies are discussed in the context of batch size and drift pattern complexity. Our drift correction tool offers a practical, simplified approach to drift correction and batch combination in large metabolomics studies.
URI: https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85029488380&origin=inward
http://cmuir.cmu.ac.th/jspui/handle/6653943832/56669
ISSN: 18733778
00219673
Appears in Collections:CMUL: Journal Articles

Files in This Item:
There are no files associated with this item.


Items in CMUIR are protected by copyright, with all rights reserved, unless otherwise indicated.